The Millennium® - the infant ventilator and pediatric ventilation system from Sechrist Industries, Inc. was designed for maximum performance and user-friendly convenience.
The Millennium® Infant Ventilator is a simple, reliable ventilation for neonatal, infant, and pediatric patient requirements. It has a unique pressure sensing device that synchronizes the patient's natural breathing cycle with the neonatal ventilator as well as a maximum sensitivity of 0.10 cmH20, and a response time of 5-40 milliseconds depending upon the sensitivity settings.
TECHNICAL SPECIFICATIONS (as mentioned on company's brochure)
Recommended Applications
Neonatal, infant& pediatric patients < style="font-weight: bold;">Flow
2-32 lpm; flush 40 lpm
Modes of Operation
AC, SIMV/IMV, CPAP, CPAP with backup ventilation, Standby
Inspiratory Time
0.1 to 3.0 seconds
I : E Ratio
Normal Range:
1:1.0 to 1:10 (sens. off)
1:1.0 to 1:99 (sens. on)
Inverse Range:1.0:1 to 4.0:1
Rate
2 to 150 breaths per minute
Expiratory Pressure
0 to 20 cmH2O
Inspiratory Pressure
5 to 70 cmH2O
Waveform
Tappered to square
Manual Breath
Electronically controlled, based upon the inspiratory pressure, inspiratory time and waveform control setting
Alarms
Alarm Delay Time
3 to 15 seconds
Alam Mute
30 Seconds
Safety Pressure Relief Valve
15 to 85 cmH20
Indicators
Dimensions
Weight
21 lbs (9.5kg)
Gas Consumption
14 lpm above the continuous flow setting
Power Requirements
This Medical Equipment will help people to improve a life.